Conference proceedings article

Modeling a safety-related system for continuous non-invasive blood pressure monitoring



Publication Details
Authors:
Sheng, H.; Schwarz, M.; Börcsök, J.
Editor:
IEEE
Publisher:
Curran Associates
Place:
Red Hook, New York

Publication year:
2013
Pages range :
254-257
Book title:
2013 IEEE Biomedical Circuits and Systems Conference (BioCAS)
DOI-Link der Erstveröffentlichung:


Abstract
In recent years, Pulse Transit Time (PTT) - based non-invasive continuous blood pressure monitoring systems have been investigated extensively. But the most relevant studies did not pay attention to the safety requirement of the system. In this paper a method is proposed to model the safety-related system for continuous noninvasive blood pressure monitoring. The V model, 1oo2 system and safe Bluetooth communication are used to enhance the safety of the system.